SafetyAlerts
December 4, 2003
Respironics Issues Voluntary Recall of ComfortGel®
Mask
Respironics, Inc. (NASDAQ/NMS Symbol: RESP) has voluntarily
recalled 5,293 ComfortGel® nasal masks distributed from September
5, 2003 through September 15, 2003. These nasal masks, which were
part of Respironics initial launch of the ComfortGel Mask, have
been recalled because the company discovered that some masks
shipped to Home Health Care Providers or distributors did not
contain a built-in exhalation port. The exhalation port is
required to allow exhaled air to properly exhaust. The masks were
shipped without the exhaust port used to exhaust all of the
exhaled CO2, which potentially could result in patients
re-breathing CO2, associated decreased oxygen levels and increased
CO2 levels in the blood. The ComfortGel nasal mask can only be
obtained through a physician prescription.
The ComfortGel masks are used in the treatment of Obstructive
Sleep Apnea (OSA) in conjunction with a Continuous Positive Airway
Pressure (CPAP) or Bi-Level Positive Airway Pressure (Bi-PAP)
device which supplies a continuous stream of air which keeps the
patient’s airway open during sleep to diminish apnea, or
awakening, episodes.
Respironics self-reported the exhalation port issue to the Food
and Drug Administration District in Philadelphia, Pennsylvania on
September 26, 2003, when the issue came to the company’s
attention.
A corrective action was begun immediately in which Respironics
contacted all customers who were shipped masks during the
September 5 to September 15 timeframe. The company has completed
contact with all customers who received shipments of the masks and
will continue inspection and replacement efforts until all
suspected masks are returned by customers. As of today, 82% of the
recalled masks have been returned to Respironics and less than 3%
have been determined to be affected.
There have been no patient injuries to date. Respironics will work
closely with the FDA to ensure that all masks in question are
returned to the company for thorough inspection and repair or
replacement. Respironics, Inc. is notifying its distributors and
customers by telephone and written communication and is arranging
for return of all recalled products. Consumers with questions may
contact the company at 1-800-345-6443. |